Feasibility of Coronary CT Angiography-based Fractional Flow Reserve Using a Clinical Photon Counting Detector CT System
Emese Zsarnóczay1, Daniel Pinos2, U. Joseph Schoepf2, Nicola Fink2, Jim O'Doherty2, Milan Vecsey-Nagy2, Pál Maurovich-Horvat1, Tilman Emrich2, Akos Varga-Szemes2
1: Medical Imaging Centre, Semmelweis University
2: Medical University of South Carolina
Poszter
Poster Session N - Cardiovascular Medicine and Research 2.
English
Cardiovascular Medicine and Research
Introduction: Coronary computed tomography angiography (CCTA)-based fractional flow reserve (CT-FFR) allows for noninvasive determination of the functional severity of anatomic lesions in patients with coronary artery disease.
Aims: The aim of this study was to intra-individually compare CT-FFR between photon-counting detector (PCD) and conventional energy-integrating detector (EID) CT systems.
Method: In this single-center prospective study, subjects who underwent clinically indicated CCTA on an EID-CT system were recruited for a research CCTA on PCD-CT within 30 days. Image reconstruction settings were matched as closely as possible between EID-CT (Bv36 kernel, iterative reconstruction strength level 3, slice thickness 0.5mm) and PCD-CT (Bv36 kernel, quantum iterative reconstruction level 3, virtual monoenergetic level 55keV, slice thickness 0.6mm). CT-FFR was measured semi-automatically using a prototype on-site machine learning algorithm by two readers. CT-FFR analysis was performed per-patient and per-vessel, and a CT-FFR≤0.75 was considered hemodynamically significant.
Results: A total of 22 patients (63.3±9.2 years; 7 women) were included. Median time between EID-CT and PCD-CT was 5.5days. Comparison of CT-FFR values showed no significant difference and strong agreement between EID-CT and PCD-CT in the per-vessel analysis (0.88 [0.74-0.94] vs. 0.87 [0.76-0.93], P=0.096, mean bias 0.02, limits of agreement [LoA] -0.14/0.19, r=0.83, ICC=0.92), and in the per-patient analysis (0.81 [0.60-0.86] vs. 0.76 [0.64-0.86], P=0.768, mean bias 0.02, LoA -0.15/0.19, r=0.90, ICC=0.93). All included patients were classified into the same category (CT-FFR>0.75 vs ≤0.75) with both CT systems.
Conclusions: CT-FFR evaluation is feasible with PCD-CT and it shows a strong agreement with EID-CT-based evaluation when images are similarly reconstructed.